Skip to main navigation Skip to search Skip to main content

Design and construction of in-house source holder for assaying the source strength of I-125 seed, model #: STM1251

  • Sarathiel Tuyizere
  • , Joseph Makuraza
  • , Samuel Nii Adu Tagoe
  • , Edem Kwabla Sosu
  • , Alhassan Mohammed Baidoo
  • Rwanda Military Referral and Teaching Hospital (RMRTH)
  • University of Rwanda
  • Adventist University of East and Central Africa
  • University of Ghana
  • Sweden Ghana Medical Centre

Research output: Contribution to journalArticlepeer-review

Abstract

Introduction: Assessment of the source strength is a mandatory task performed by medical physicists to ensure accurate, optimal dose calculation and delivery to a patient undergoing brachytherapy (BT) treatment. Furthermore, there is the need to verify source strength provided by a vendor of BT sources before their clinical use to ensure traceability to international standards such as the American Association of Physicists in Medicine and European Society for Therapeutic Radiology and Oncology (AAPM-ESTRO) recommendations. Material and methods: An in-house source holder that could be accommodated within the well-type ionisation chamber (WTIC) was designed and constructed from Perspex (PMMA). The constructed source holder was designed to hold a cartridge of I-125 BT seed within the WTIC to facilitate batch source strength measurement. The source holder ensuring reproducibility of I-125 seed placements during source calibration, also enabled placement of a cartridge such that the point of concentration of activity of I-125 seeds coincided with the calibration point. Results: The mean percentage deviation of the source strengths was (1.155 ± 0.854) % which agrees with AAPM-ESTRO recommendations. For all batches of I-125 seeds investigated in this work, we found that the WTIC response was linear. Conclusions: We, therefore, recommend that the methodology adopted to assay the source strength of Bard I-125 seed (model #: STM1251) be used by the BT departments offering low dose rate permanent interstitial BT services.

Original languageEnglish
Pages (from-to)118-124
Number of pages7
JournalPolish Journal of Medical Physics and Engineering
Volume32
Issue number2
DOIs
Publication statusPublished - 1 Jun 2026

Keywords

  • AAPM-ESTRO
  • Calibration
  • I-125 seeds
  • Source holder
  • WTIC

Fingerprint

Dive into the research topics of 'Design and construction of in-house source holder for assaying the source strength of I-125 seed, model #: STM1251'. Together they form a unique fingerprint.

Cite this